الوصف
The Yokogawa ISC40 is an inductive, toroidal conductivity sensor whose encapsulated coils do not contact the process liquid. This arrangement helps avoid electrode polarization and supports higher-conductivity or fouling-prone measurements when the process and material selection are suitable.
Verified technical overview
| مبدأ القياس | Inductive (toroidal/electrodeless) conductivity |
|---|---|
| Published measurement coverage | Approximately 50 to 2,000,000 µS/cm; low-end suitability must be checked |
| Sensor materials | PEEK variants for general service and PFA/Teflon variants for selected aggressive-media duties |
| التثبيت | Holders and adapters support in-line, off-line, insertion and selected sanitary-flange arrangements |
| Selection constraint | Chemical compatibility, temperature, pressure, concentration curve and analyzer compatibility are order-code-dependent |
Engineering selection checklist
Define conductivity or concentration range, chemical composition, temperature, pressure, solids, coating tendency and fitting style. Select PEEK or PFA construction only after a compatibility review and confirm the analyzer, holder and extension-cable requirements.
| Process data | Measurement objective, normal/minimum/maximum value, medium chemistry, temperature and pressure |
|---|---|
| التثبيت | Sample or immersion arrangement, connection, cable distance, cleaning and maintenance access |
| System interface | Compatible analyzer or host, power, output, communication and alarm requirements |
| Documentation | Confirm the current Yokogawa general specification and complete order code before purchase |
Industry application guidance
Typical uses include chemical concentration, brine, scrubber, boiler blowdown and clean-in-place return monitoring. For CIP service, the conductivity-to-concentration relationship and temperature compensation must be validated for the actual cleaning chemistry.
Related products and engineering resources
- أجهزة التحليل
- Beverage Analysis Instrumentation Guide
- محلل السائل بأربعة أسلاك Yokogawa FLXA402
- محلل السائل بسلكين Yokogawa FLXA202
- Pharmaceutical & Biotechnology Solutions
Official manufacturer documentation
Official Yokogawa product page · Official Yokogawa general specification or product data
Published values above are tied to the stated Yokogawa model family and may vary by order code. Confirm materials, ranges, process limits, approvals and accessories for the exact configuration. No availability, price, project or performance claim is implied.
Request a quotation
Send the model, process conditions and required configuration to sales@santanfluid.com, WhatsApp China +86 136 9022 3185 or WhatsApp Indonesia +62 823 1767 3750.
ملخص المواصفات الفنية القياسية
استخدم هذه الحقول للاختيار الأولي. تعتمد القيم النهائية على رمز نموذج المصنع الكامل والخيارات المحددة.
| النموذج | yokogawa-isc40-inductive-conductivity-sensor |
|---|---|
| نطاق القياس | يعتمد على التكوين؛ أكد النطاق المطلوب والكود النموذج الكامل. |
| الدقة | يعتمد على التكوين؛ أكد فئة الدقة والنطاق المعاير. |
| توصيلة العملية | يعتمد على التكوين؛ أكد الحجم والتصنيف والمعيار والمواد المبللة. |
| إشارة الخرج | يعتمد على التكوين؛ تأكيد متطلبات التناظري والرقمي والاتصالات. |
| الشهادات | أكد موافقات المناطق الخطرة والسلامة الوظيفية والصحية والبحرية أو نقل الحضانة برمز النموذج. |
| صناعات التطبيق | يعتمد على التطبيق؛ أكد الوسط وظروف العملية ومتطلبات الصناعة. |

التقييمات